KODAK In-Vivo Imaging System FX Pro Specifications
|
Camera
|
|
CCD |
2048 x 2048 pixels |
Cooling |
-29°C absolute |
Lens |
Computer controlled 10X zoom, 20 - 200mm f2.8 |
|
Illumination
|
|
Fluorescence |
Automated 15-position selectable multiwavelength epi-illumination, 380 - 780 nm – includes 14 excitation filters and 175 watt xenon illumination source |
|
UV (306 nm) transillumination (optional) |
White Light |
White light epi-illumination and white light transillumination |
|
Emission Filters
|
Automated selectable wide angle filters, 440 - 830 nm - includes 4 emission filters |
|
Performance
|
|
Field of View |
2 x 2 cm to 20 x 20 cm, continuous automated zoom |
Optical Resolution (max.) |
|
Data Acquisition |
16-bit single capture
n-bit data acquisition |
Binning |
1x2, 2x2, 1x4, 2x4, 4x4, 1x8, 2x8, 4x8, 8x8, 16x16 |
| Dark Current |
.003 e-Pixel/Sec |
Read Noise |
9 electrons (Nominal) |
Dynamic Range |
>4 OD |
Quantum Efficiency |
@ 470nm > 55% |
|
Detection Modes
|
Luminescence
Multiwavelength Fluorescence
Absorbance
Radioisotopic
X-Ray |
|
Exposure Modes
|
Single Capture
Multiple Capture
Progressive Exposure
Time Lapse Exposure |
|
Computer
|
|
Interface |
IEEE1394 (FIREWIRE) |
Operating Systems |
WINDOWS 2000/XP
MACINTOSH OS |
| Digital X-ray |
|
Energy Range |
Approximately 12-35 kVP |
Net Weight |
160 lbs |
Maximum Current |
Approximately 150uA |
Spot Size |
< 50 U |
Target Material |
Tungsten |
Window Filtration |
Beryllium |
Cone of Illumination |
>33 degrees |
Field of Illumination |
20 X 20 |
Instrument |
|
Approximate Dimensions |
(in) 69 x 24 x 52 (cm) 175 x 66 x 122 (l x w x h) |
Approximate Net Weight |
226 lbs |
Power Requirements |
120VAC, 7A
230VAC, 3.5A
|
|
| |
|
|
|
|
|
|
|
|
|
|
|